Uncertainty & measurement

Repeated measurements & timing

Summarize repeated readings and timing uncertainty.

Your measurements

Number formats & units

Use decimal dots in datasets. Single values also accept decimal commas and scientific notation such as 1.2e-3 or 1.2 × 10⁻³.

Repeated Measurement Uncertainty & Timing Calculator inputs
Same unit as each total reading. Use a justified uncertainty from the apparatus or task.
Use 1 for ordinary repeated measurements.

Calculated locally. No account or uploads.

Result & working

Example

The calculator is loading. You can read the worked example below.

Copy & export

Formula & example

The relationship

xˉ=∑ixin\bar x=\frac{\sum_i x_i}{n}range=xmax⁡−xmin⁡\text{range}=x_{\max}-x_{\min}
Δx=max⁡ ⁣(range2, supplied uncertainty floor)\Delta x=\max\!\left(\frac{\text{range}}2,\ \text{supplied uncertainty floor}\right)
T=xˉNT=\frac{\bar x}{N}ΔT=ΔxN\Delta T=\frac{\Delta x}{N}p=100ΔT∣T∣p=\frac{100\Delta T}{|T|}
What the variables mean
SymbolMeaning
nnNumber of repeated readings.
NNExact cycle/item count per reading, a positive integer.
FloorAn explicitly supplied lower bound for uncertainty. It is not estimated from the data.

Worked example

Repeated readings / total times
20.1, 20.3, 20.2, 20.0
Absolute instrument / method uncertainty floor
0.1
Exact number of cycles or items per reading
10
Reading and result unit
s

(2.02 ± 0.02) s

Follow the example steps
  1. Mean = 80.6 / 4 = 20.15 s.
  2. Range = 20.3 − 20 = 0.3; half-range = 0.15 s.
  3. Chosen uncertainty = max(0.15, 0.1) = 0.15 s.
  4. Value per cycle/item = 20.15/10 = 2.015; uncertainty = 0.15/10 = 0.015 s.
  5. Final reported value: (2.02 ± 0.02) s.

Common mistakes

Half-range and standard error are different methods.

Incorrect approach

9, 10, 11: Δx = 1/√3

Correct approach

9, 10, 11: Δx = (11 − 9)/2 = 1

More checks before you submit
  • Do not divide the half-range by √n; that produces neither the half-range nor sample SEM.
  • Distinguish repeats n from cycles per timing N.
  • Do not remove an extreme value without investigating and explaining it.
Method notes

Use half the range when that is the convention requested in a school experiment. This tool applies the larger of half-range and the supplied floor; this is a stated reporting convention, not an IB rule for every instrument or a universal statistical formula.

Timing several cycles can reduce the relative effect of a fixed start/stop uncertainty. Count complete oscillations and keep the cycle count exact. The tool also displays sample SD and SEM for comparison, without substituting them for the half-range.

Identical repeated readings do not prove zero uncertainty. Resolution, calibration and method limits still matter. A systematic offset is not removed by repetition.