Experimental data & uncertainty · SL/HL

Calibration curve

Fit standards and infer an unknown from its response.

Your values

Number formats & units

Use decimal dots in datasets. Single values also accept decimal commas and scientific notation such as 1.2e-3 or 1.2 × 10⁻³.

Calibration Curve & Unknown Concentration Calculator inputs

Calculated locally. No account or input upload.

Result & working

Example

The calculator is loading. The worked example is also shown below.

Copy & export

Formula & example

The relationship

y=mx+cy=mx+c
xunknown=yunknown−cmx_{\mathrm{unknown}}=\frac{y_{\mathrm{unknown}}-c}m

R2R^{2} evaluates goodness of fit to the chosen linear trend

What the variables mean
SymbolMeaning
xxKnown standard concentration.
yyInstrument response such as absorbance when a linear calibration is justified.

Worked example

Standard concentrations x
0.00 0.20 0.40 0.60 0.80 1.00
Measured responses y
0.02 0.18 0.35 0.53 0.69 0.87
Unknown response y
0.44
Concentration unit
mol dm⁻³
Response unit
absorbance

0.5 mol dm⁻³

Follow the example steps
  1. OLS fit: y = 0.8514286x + 0.01428571
  2. R² = 0.9997073
  3. x_unknown = (y_unknown − c)/m = (0.44 − (0.01428571)) / 0.8514286 = 0.5 mol dm⁻³

Common mistakes

Use the intercept when converting a response to a concentration.

Incorrect approach

y = 2x + 1, y = 5: x = 5/2 = 2.5

Correct approach

x = (5 − 1)/2 = 2

More checks before you submit
  • A calibration curve should use standards prepared and measured under conditions appropriate to the method.
Method notes

Check whether the unknown response lies within the calibrated range. Extrapolation beyond standards is less defensible and should be identified. A high R² alone does not prove the model or eliminate systematic error.